Maintenance 127
Performing the startup wash (WASH3)
Materials required:
• Deionized water
Time: 26 minutes
Analyzer mode: READY
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
The daily startup wash rinses the probe lines, reaction cuvettes and dilution cuvettes.
Location Position Wash Solution
1
CTT-15 ISE Detergent Solution
1
CTT-16 Deionized water
1
CTT-49 10% Cuvette Wash Solution (Daily)
5% Reagent Probe Wash 3 (Weekly)
1
CTT-50 Deionized water
1
CTT-51 Deionized water
2
RTT1-53 Reagent Probe Wash 1
2
RTT1-54 Reagent Probe Wash 2
2
RTT1-55 10% Cuvette Wash Solution (Daily)
5% Reagent Probe Wash 3 (Weekly)
2
RTT1-56 Deionized water
3
RTT2-53 Reagent Probe Wash 1
3
RTT2-54 Reagent Probe Wash 2
3
RTT2-55 10% Cuvette Wash Solution (Daily)
5% Reagent Probe Wash 3 (Weekly)
3
RTT2-56 Deionized water
Figure 5-12. Location of wash solutions on the CTT and RTT trays