130 ADVIA 1800 Operator’s Guide
NOTE: Do not perform the ISE electrode wash more than 3 times per day (once as
part of the shutdown wash and twice on a per-shift basis). Pour fresh ISE Detergent
into a cup, not a tube, before each wash from the CTT.
1. At the Menu Panel, select Maint., then select ISE Operation.
2. At the ISE Operation window, in the Period.wash area, select OFF, then select
Set.
3. In the Wash Electrode area do the following:
a. Place a cup of the ISE Detergent in the identified position on the screen.
b. In the Container field, select the type of container for the wash solution.
The recommended type of container is 6 : 2mlCUP/Adp.
4. In the Wash Electrode area, select Execute, then select Yes when prompted.
5. At the ISE Operation window, in the Period.wash area, select ON, then select
Set.
6. Close the window, then select
Yes when prompted.
Recording ISE slopes
Once a day, record the slopes from a successful ISE calibration on the Maintenance Log.
The slopes are provided on the ISE Monitor, RBL/Calibration History, and RealTime
Monitor windows following a successful calibration.