Maintenance 161
1 Terminal 2
2 Joint Connector
3 Joint Holder
4
Phillips Screws (4 places)
5 Probe
6 Terminal 1
7 Flange
Figure 5-29. Probe components
5. Cover the 12-volt LLS amplifier and wires with a lint-free paper towel.
6. Using the pliers, loosen the joint connector (
2) turning counterclockwise, then
unfasten and remove it by hand.
7. Loosen but do
not remove the 4 Phillips screws (4).
NOTE: Be careful not to drip any fluid from the probe onto the amplifier. If needed,
drain fluid into a container.