58 ADVIA 1800 Operator’s Guide
In this window, you can delete specific reagent codes and specific reagent barcodes from
the ADVIA reagent barcode database. Only use this function to delete an old reagent code
or reagent barcode that may have the same lot number and container number as a new
reagent container you want to place on the system. In most circumstances, you do not have
to use this functionality.
The barcode maintenance database can store a maximum of 200 reagent codes (assays)
and up to 2000 individual reagent barcodes. If the database exceeds the limit of 2000
barcodes, the oldest stored barcode is replaced with the most current barcode.
The reagents are listed in the same order as they display in the System Test List window.
The Reagent Barcode and Date Opened on System information for one or more reagents
displays when a Reagent Code is selected.
Perform any of the following tasks at this window:
clear all
clear reagent code(s)
clear reagent barcode(s)
clear specific date
clear range of dates
QC windows
Select the QC button to display the following QC menu.
ADVIA QC window
ADVIA QC provides both real-time and long term evaluation of analyzer and method
performance.
To automatically transfer data to this QC manager, at the System Monitor window, in the
ADVIA QC Transfer area, select
Yes.
Perform any of the following tasks at this window:
collect control results
calculate and display statistical data
assess data errors
identify QC violations
review control results
create printed reports
identify and report events such as lot changes and calibration dates
Real-time QC window
Use this window to review the performance for two controls simultaneously using a
Levey-Jennings chart (x-chart) or a twin chart.