96 ADVIA 1800 Operator’s Guide
Examples of shutdown settings
The following examples demonstrate some common scenarios that you can implement
with the Shutdown Set feature of the System Startup/Shutdown Settings window.
System available during the night
To have both the computer and analyzer available for night emergencies, use the
following:
System p.s. Power OFF check box is cleared (no check mark).
PRIME
Select
.
Do check box is cleared (no check mark).
WASH2 Do check box is selected (check mark present).
Do check box is cleared (no check mark).
Automatic startup from the Sleep state
In the Sleep state, the Power Panel Operate/Standby switch (1) remains set to OPERATE.
However, electrical power is supplied only to the necessary components, and the Power
indicator (
2) is off.
To place system in the SLEEP state, use the following:
Power OFF check box is selected (check mark present).
System
Select
.
WASH1 Do check box is cleared (no check mark).
Do check box is selected (check mark present).
WASH3 Do check box is cleared (no check mark).
Automatic startup with analyzer power kept ON
To keep analyzer power on until the automatic start up, use the following:
Power OFF check box is cleared (no check mark).
PRIME
Select
.
Do check box is cleared (no check mark).
WASH2 Do check box is selected (check mark present).
Do check box is cleared (no check mark).
NOTE: If System is set to Not do, you must select System end in the System list before
the automatic startup setting time occurs. Otherwise, the automatic startup does not occur.
In this case, you cannot shut down Windows even if you select
Shutdown on the
[ADVIA] Startup window. Additionally, the CRT screen saver does not operate.