Examination
Cassette Exposures in the Spotfilm Device (optional)
18 / 52 AXD3-340.620.13.01.02 Operator Manual
Selection of the dominants
The dominants are filed in the organ programs or they can be set on the integrat-
ed generator control console.
Multiple selection of the dominants is possible.
◆ Press one of the keys.
Left dominant
Middle dominant
Right dominant
The position of the measuring field/of the measuring field in relation to the object
can be checked on the LIH image by means of brief fluoroscopy. A diagram with
the position and size of the relevant measuring field is displayed briefly on the
monitor.
Warning
Please make sure when changing the selection of the dominants that the
correct dominant is always selected.
If a selected left or right dominant becomes invalid temporarily, e.g. due to colli-
mation, then the middle dominant is selected and displayed automatically. The
deselected dominant is displayed crossed through.
The originally selected dominant is active and displayed again only at the 1st
pressure point of the exposure release button (precontact) or when the collima-
tor is opened again.