Examination
Cassette Exposures in the Spotfilm Device (optional)
AXIOM Iconos R200 AXD3-340.620.13.01.02 19 / 52
Releasing the exposure
Release the exposure with the exposure release button in the system remote
control console or via the foot switch in the control room, or table-side (op-
tion). The foot switch has no precontact.
The signal lamps for radiation must light up briefly.
Organ programs
All parameters necessary for the exposure are stored in organ programs.
Refer to the POLYDOROS SX register for further details on the organ
programs and their programming.
Single exposures / serialography
Single exposure
Single exposure selected: the LED does not light up
Select a suitable segmentation program. Possible selection:
Full format
Segmentation 2 on 1 / 3 on 1 / 4 on 1
If you now release the exposure, you receive one single exposure per exposure
release.
Warning
If these radiation ON indicators light up although no switch has been actuated
and the current fluoroscopic image cannot temporarily be seen on the monitor:
Switch the system off immediately!
Notify the SIEMENS Uptime Service.
Discontinue operation of the system.