Examination
Digital Radiography
AXIOM Iconos R200 AXD3-340.620.13.01.02 25 / 52
Spotfilming without tomography
Selection
◆ Press this key.
– Digital radiography is selected: the LED lights up
◆ At the integrated generator control console, select an organ program for
– Single image or
– Serialography or
–DSA
Single image - serialography
Not possible in DSA.
The display is in the operating area of the live image monitor.
Changing selection
◆ If you press this key
– the system switches over to single image in the “Serialography organ pro-
gram”
– single image remains selected in the “Single image organ program”
Frame rate
Operation is on the infrared control of the FLUOROSPOT Compact.
◆ During a native series the programmed frrame rate can be rduced in a maxi-
mum of 2 stages by pressing the button, if the corresponding ruductions have
been programmed in the relevant organ program (see FLUOROSPOT Com-
pact).
DR
The image intensifier format cannot be changed during an exposure series. Only
the collimation is corrected on changing over to a larger zoom level.
A changed organ program for serialography remains retained only until you se-
lect another / same organ program or switch the system off.
F/s