Siemens AXIOM Iconos R200 - Spotfilming Without Tomography; Selection; Single Image - Serialography; Changing Selection

Siemens AXIOM Iconos R200
238 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Examination
Digital Radiography
AXIOM Iconos R200 AXD3-340.620.13.01.02 25 / 52
Spotfilming without tomography
Selection
Press this key.
Digital radiography is selected: the LED lights up
At the integrated generator control console, select an organ program for
Single image or
Serialography or
–DSA
Single image - serialography
Not possible in DSA.
The display is in the operating area of the live image monitor.
Changing selection
If you press this key
the system switches over to single image in the “Serialography organ pro-
gram”
single image remains selected in the “Single image organ program”
Frame rate
Operation is on the infrared control of the FLUOROSPOT Compact.
During a native series the programmed frrame rate can be rduced in a maxi-
mum of 2 stages by pressing the button, if the corresponding ruductions have
been programmed in the relevant organ program (see FLUOROSPOT Com-
pact).
DR
DR
The image intensifier format cannot be changed during an exposure series. Only
the collimation is corrected on changing over to a larger zoom level.
A changed organ program for serialography remains retained only until you se-
lect another / same organ program or switch the system off.
F/s