Examination
Digital Radiography
AXIOM Iconos R200 AXD3-340.620.13.01.02 29 / 52
Releasing the exposure
❏ To release the exposure, actuate the exposure release button in the unit re-
mote control console.
❏ The radiation ON indicators must light up during the exposure.
Radiation interruption If you let go of the exposure release button during radiation, the exposure is in-
terrupted immediately and the tomographic movement is stopped.
Exposure end Letting go of the exposure release button or elapsing of the tomographic time
ends the exposure.
Organ programs
All parameters necessary for the exposure are stored in organ programs.
➩ Refer to the FLUOROSPOT Compact and POLYDOROS SX operating instruc-
tions for further details on the organ programs and their programming.
Programs
The following programs can be stored:
– Tomography organ programs
Creation of programs
➩ Refer to the FLUOROSPOT Compact operating instructions.
Temporary changes of programs
For the following parameters stored in the organ program, it is possible to change
them temporarily with the corresponding operating elements:
– Film density
– Cu prefilter
Warning
If these radiation ON indicators light up although no switch has been actuated
and the current fluoroscopic image cannot be seen on the monitor:
◆ Switch the generator off immediately!
◆ Notify the SIEMENS Uptime Service.
◆ Discontinue operation of the system.
The changed parameters apply only until you select another / same organ pro-
gram or switch the system off.