Examination
Digital Radiography
AXIOM Iconos R200 AXD3-340.620.13.01.02 31 / 52
Releasing the exposure
Release the exposure with the exposure release button on the system remote
control console or via the foot switch in the control room, or table-side (op-
tion).
Radiation is released long as long as you continue to press the exposure re-
lease button.
The radiation ON indicators must light up.
Radiation interruption If you let go of the exposure release button during radiation, the exposure is in-
terrupted immediately.
Exposure end Letting go of the exposure release button , a full image memory or reaching the
programmed maximum scene time ends the exposure series.
Examination procedure
Preparation
Prepare the digital imaging system
Move the tube assembly into the 0° position
Position the patient
Set a nominal source-image intensifier distance of 115 cm or 150 cm
Bring the X-ray system and tabletop into the intended starting position.
You must ensure that the travel range for the X-ray system is sufficient.
Press this key.
The LED lights up: Periscanning is selected.
Serialography is selected.
Warning
If these radiation ON indicators light up although no switch has been actuated
and the current fluoroscopic image cannot be seen on the monitor:
Switch the system off immediately!
Notify the SIEMENS Uptime Service.
Discontinue operation of the system.
DR
DR
The LEDs in the Periscanning and start keys flash if the conditions for exposure
are not fulfilled.