Examination
Digital Radiography
AXIOM Iconos R200 AXD3-340.620.13.01.02 35 / 52
Programming run
Proceeding from the abdomen (starting position), the primary collimators, semi-
transparent wedge filters and dominant, if required, are set step by step (up to
the last position).
You must deflect the joystick or press the key for “X-ray system longitudinal
movement” in the movement direction for each step as long as the next step po-
sition is reached. Once the next step position is reached, the X-ray system stops
automatically and the corresponding position number appears on the remote con-
trol console.
The last set collimator positions, filter positions and dominants are stored auto-
matically assigned to each step position.
A maximum of 6 positions (5 steps) is possible.
After the programming run is completed, the X-ray system must be moved back
to the starting position (abdomen) (automatic stop).
Filling run
Initiate the individual steps of the X-ray system using the joystick or the key for
“X-ray system longitudinal movement” corresponding to the contrast medium
flow with monitoring on the live image monitor.
Selection
◆ Press this key.
– Peristepping is selected: the LED lights up.
Deselection
◆ Press this key or select another exposure system.
– Peristepping is deselected.
DR
There is no automatic deselection of Peristepping after the end of the scene.
DR