Examination
Digital Radiography
AXIOM Iconos R200 AXD3-340.620.13.01.02 35 / 52
Programming run
Proceeding from the abdomen (starting position), the primary collimators, semi-
transparent wedge filters and dominant, if required, are set step by step (up to
the last position).
You must deflect the joystick or press the key for “X-ray system longitudinal
movement” in the movement direction for each step as long as the next step po-
sition is reached. Once the next step position is reached, the X-ray system stops
automatically and the corresponding position number appears on the remote con-
trol console.
The last set collimator positions, filter positions and dominants are stored auto-
matically assigned to each step position.
A maximum of 6 positions (5 steps) is possible.
After the programming run is completed, the X-ray system must be moved back
to the starting position (abdomen) (automatic stop).
Filling run
Initiate the individual steps of the X-ray system using the joystick or the key for
“X-ray system longitudinal movement” corresponding to the contrast medium
flow with monitoring on the live image monitor.
Selection
Press this key.
Peristepping is selected: the LED lights up.
Deselection
Press this key or select another exposure system.
Peristepping is deselected.
DR
DR
There is no automatic deselection of Peristepping after the end of the scene.
DR
DR