Examination
Digital Radiography
AXIOM Iconos R200 AXD3-340.620.13.01.02 37 / 52
If the LED in the start key flashes,
◆ press the key.
– The compression device moves into park position, if present
– Oblique setting moves into 0° position
– A source-image intensifier distance of 115 cm is set
Once these positions are reached, the flashing LED goes out.
◆ Select an organ program for Peristepping.
◆ Set the following parameters with the corresponding operating elements:
– Primary rectangular collimators
– Cu prefilter
– Semitransparent wedge filters
Adapting the
dominant
◆ Press one/several keys
– The dominant preset in the organ program is overwritten.
Prepare the injector ◆ Connect the injector plug to the electronics cabinet (only with radiation-syn-
chronous injector release (option)).
◆ Set the injection data and switch to ready for injections.
The LEDs in the Peristepping and start keys flash if the conditions for exposure
are not fulfilled.
Start
Zoom formats cannot be selected. The I.I. full format of 40 cm is always select-
ed.
The primary collimation is not set automatically to the full format, but must be
adapted according to the object.
If a left / right dominant or both dominants are selected, the system switches
over automatically to the central circular dominant if the preselected dominant is
covered by the filters, since this would inevitably result in a faulty exposure.