Examination
Digital Radiography
42 / 52 AXD3-340.620.13.01.02 Operator Manual
Examination procedure for abdominal survey images
Preparation
Prepare the digital imaging system.
Select the DR system at the unit remote control console.
Move the patient table into the 0° position (horizontal).
Position the patient, head on the left.
For reasons relating to the exposure technology, the scanning direction for DR
scanning should be from the pelvis headwards.
Set vertical projection.
Set the largest possible SID.
Perform the examination, e.g. colon - double contrast, as usual.
Press this key.
The LED in the start key lights up if the conditions for scanning are not ful-
filled.
DR scanning is selected.
I.I. full format is selected.
The rectangular height diaphragm moves automatically to the configured
value (approx.10 - 12 cm)
The display shows 3 scanning programs with 2 directions each.
The selected scanning program is highlighted.
By pressing the key again the next possible scanning program with direction be-
comes active.
Select the scanning program for abdomen
left arrow: scanning from right to left
right arrow: scanning from left to right
DR
Ln
S
i
pne<- -
gelg-->o<- -
<- - Abdomen - ->
-->