POLYDOROS SX 65/80
Integrated Control Console
AXIOM Iconos R200 AXD3-340.620.13.01.02 19 / 36
❏ Recommended characteristic curves for organ programs in 0-point technique
– These suggestions correspond to the SIEMENS basic exposure table,
whereby only the part relevant for the 0-point technique has been listed.
kV
60
66
73
81
90
102
0
10 20 30 40
117
133
cm Wasserwert
WW
50
C60
C61
C62
C63
150
Water value
(WV)
Organ kV Characteristic curve
Cranium:
Survey of cranium p.-a., a.-p
Cranium lateral
Cranium axial
Petrosal bones sagittal
Petrosal bones Stenvers
Paranasal sinuses p.-a.
Optical nerve cavity Rhese
70
66
85
70
70
70
66
C 14
C 13
C 18
C 15
C 14
C 14
C 13