Accessories
Optional Accessories
AXIOM Iconos R200 AXD3-340.620.13.01.02 25 / 54
Removal ◆ Move the patient table to the horizontal position.
◆ Move locking handle on the rear of the footboard in arrow direction and hold
fast.
◆ Pull the metal clip of the foot restraint out of the hole in the footboard.
◆ Loosen the straps.
◆ Lift the patient’s feet out of the foot restraint.
Control/care note Check the foot restraints for correct condition at regular intervals of approx. 2
months.
Check especially the firm seating of the fastening rivets, of supporting belts and
closing belts, as well as the completeness of their seams and the condition of the
leather for cracks and porous places.
Before using the foot restraint, always check that the safety device (3) is present.
If this safety device is not present, then the foot restraint must not be used any
more.
In case of defects to the foot restraint take this out of service and obtain a re-
placement through the Siemens Uptime Service.
We recommend using a commercially available impregnating pump spray for
cleaning and care.