30 / 54 AXD3-340.620.13.01.02 Operator Manual
Attachment Move the patient table to the horizontal position.
Position the patient.
Loosen the cap screw (1).
Place the cassette holder on the accessory rail and move it to the exposure
position.
Tighten the cap screw (1).
Push and pull the cassette holder to ensure that it is seated properly.
Lift the tensioning device (4).
Place the cassette on the lower holder.
Slide the tensioning device downward.
Ensure that the cassette is seated properly in the holder.
Loosen the cap screw (2) to permit adjustment of the height and the cap
screw (3) to permit lateral adjustment.
With the patient positioned for the examination, adjust the height and lateral
tilt of the cassette.
Tighten the cap screws (2) and (3).
Ensure that the patient is stabilized and comfortable.
4
3
2
1
Have the patient place his/her hands on the grip protection strip, the grip strip,
or the handgrip, depending on the examination procedure. This will increase the
patient’s feeling of security during the examination.