Accessories
Optional Accessories
AXIOM Iconos R200 AXD3-340.620.13.01.02 41 / 54
Attachment Move the patient table to the horizontal or vertical position as needed;
If the table is not in the horizontal position, insert the BABIX holder into the
accessory rail, position it and secure it;
Push and pull on the BABIX holder to ensure that it is seated properly;
Bring the BABIX cradle with the secured infant into the examination room;
Hang the BABIX-U cradle on the hook of the holder and position it for the ex-
amination.
Place the BABIX-Flat cradle on the horizontal patient table in the examination
position.
Removal Remove the BABIX-U from the holder;
Take the BABIX cradle with the infant out of the examination room and attend
to the infant;
Hang the BABIX cradle on the corresponding BABIX rack and store appropri-
ately.
Radiation protection
value:
The attenuation equivalent of the BABIX cradle is 1.6 mm Al.
Permissible load of the
BABIX cradle:
Size 600 mm: 6 kg
Size 700/800 mm: 10 kg
Size 1000 mm: 15 kg