AXIOM Iconos R200 AXD3-340.620.13.01.02 47 / 54
Radiation protection for tableside
examinations
Use Additional radiation protection in tableside examinations, e.g. interventional radi-
ography.
Fastening Recommendation: Set the examination unit horizontal and lower the table to
minimum height.
Push the bar (1) into the upper groove of the accessory rails up to the stop.
When in position in the stop, a spring lever in the left profile rail engages in
a notch of the bar.
Check for the firm seating of the bar on the collimator.
Warning
Use this radiation protection only with the tabletop practically horizontal up to
max. +/- 10
o
. At a greater tilt there is a danger that the lead-rubber aprons will
slip off and drop on the sterile covered patient.
1