Siemens AXIOM Iconos R200 - Stray Radiation in the Main Operating Area According to DIN en 60601-1-3

Siemens AXIOM Iconos R200
238 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
System Overview
Protective Measures
14 / 44 AXD3-340.620.13.01.02 Operator Manual
Stray radiation in the main operating area
according to DIN EN 60601-1-3
Tolerance of the air kerma measurements ± 5%
Characteristic A and C: continuous fluoroscopy 63kV, 0.8mA (antiisowatt)
Characteristic B and D: continuous fluoroscopy 110kV, 3mA
Patient table horizontal
Patient table vertical
0123456
0
50
100
150
200
cm
mGy/h
Height above floor
Air kerma
A
B
012
0
50
100
150
200
cm
mGy/h
Height above floor
Air kerma
C
D