Technical Description
Technical Data
AXIOM Iconos R200 AXD3-340.620.13.01.02 9 / 12
TV frame rate max. 25 f/s; with CAREVISION 12,5; 8; 3 pulses/s
Brightness control Constant image brightness through automatic gain control
TV standard 72 Hz progressive scan, independent of frequency
VIDEOMED DH
High-resolution TV camera with maintenance-free 1K CCD sensor for digital
fluoroscopy, fluororadiography and DSA
Dynamic response max. 69 dB, optimized for DSA
TV matrix Matrix 1024²
TV frame rate max. 30 f/s; with CAREVISION 15; 7,5; 3 pulses/s
Brightness control Constant image brightness through automatic gain control
TV standard 72 Hz progressive scanning, independent of frequency
Monitor SIMOMED HM; high-line with flicker-free display 72 Hz; 44 cm or.. 54 cm screen
diagonal;
SIMOMED HC; 1024 lines, flicker-free display 72 Hz; 44 cm screen diagonal;
LCD monitor 18”; 46 cm screen diagonal;
Tube assembly OPTITOP 150/40/80HC-100, cover with fan.
Focal spot nominal value
(ICE 60336):
0.6/1.0
Nominal power (IEC
60613):
40 kW/80 kW
Cold power: 52 kW/103 kW
Nominal voltage (IEC
60613):
150 kV
Anode angle: 12°
Anode heat storage ca-
pacity:
580 000 J (783 000 HU)
Anode rotation frequen-
cy:
Radiography: 150 Hz/ 180 Hz
Fluoroscopy: 20 to 30 Hz
Mean continuous pow-
er:
max. 450 W
Total filtration (IEC
60601-1-3):
2.5 mm Al at 80 kV