Technical Description
Technical Data
10 / 12 AXD3-340.620.13.01.02 Operator Manual
Inherent filtration: 1.5 mm Al at 80 kV
Leakage radiation (IEC
60601-1-3):
<0.8 mSv/h at 150 kV/450 W in 1 m distance
Load curves: Refer to the X-ray tube assembly operating instructions
Primary collimator
Max. field size 35 cm x 35 cm at 0.7 m SID
43 cm x 43 cm at 1.0 m SID
42.5 cm x 42.5 cm at 1.0 m SID (for collimator with DSA module)
Min. field size < 3.0 cm x 3.0 cm at 1.0 m SID
Angle of rotation ± 50° about the central beam axis
Halogen lamp Only original Siemens replacement lamps may be used!
24 V / 150 W / part no.: 8375545 G2107
Inherent filtration 1.0 mm Al at 70 kV
Additional filters 0.1 mm / 0.2 mm / 0.3 mm Cu (manual filter selection)
Radiation protection depending on regulations, up to max. tube voltage of 150 kV
Brightness of full-field
light localizer
160 lux (measured at a distance of 1m)
(if original Siemens three-field templates and dose measurement chambers are
used)
Dimensions
(height x width x depth)
183 mm x 232 mm x 352 mm (without DSA module)
262 mm x 233 mm x 380 mm (with DSA module)
Max. weight
(without accessories)
10.5 kg to 15.5 kg (with DSA module)
Max. weight
of the accessories
7kg
Connected load 30 - 40 V AC ; 50/60 Hz; 8A
Aperture angle 28° / 28°