AXIOM Iconos R200 AXD3-340.620.13.01.02 21 / 44
Do not stand at any place on the spotfilm device or on the covers of the table
support outside the marked areas provided for this.
The covers can be deformed.
Components located underneath them are damaged and thus lead to oper-
ating disturbances.
Place no objects on the operating areas of the control consoles and the table-
side control.
In vertical table positions do not use the stand column, the tube assembly sup-
port arm, the tube assembly cover or the primary collimator as seat or support.
This unallowed loading can lead to material breakage and damage to bear-
ings.
Never put contrast medium cups or open containers with liquid or pasty con-
tents on the unit, on the remote console or on the control cabinets.
Contrast medium can spill, leak or overflow into system parts and lead to
operational disturbances of the unit or to misinterpretation of exposures.
When storing contrast medium in the cup holder on the compression carriage,
use only cups with a maximum volume of 0.25 liter made of unbreakable ma-
terials, i.e. under no circumstances glass or porcelain.
Remove contrast medium traces immediately!
Measures for avoiding unwanted radiation
Before starting system movements make sure that the foot switch for fluo-
roscopy and radiography (optional) in the examination room is not in the travel
range of the image intensifier light distribution system.
Caution
Unintentional activation of control elements for movements
Collision with patient, operator or equipment
Do not load the remote console with any objects, accessories, folders or
documents.
To avoid unintentional activation of control elements for movements concerning
bed exposure, please adhere to the following workflow:
Select aquisition mode Bed exposure at the generator control console.
Tilt the system and move it to the correct SID.
Rotate the tube assembly accordingly.
Position the patient.
Control patient and system.
Don’t let patient stay in system area during absence of operator.