System Overview
Protective Measures
22 / 44 AXD3-340.620.13.01.02 Operator Manual
Safety-relevant parts subject to wear
This system contains no safety-relevant parts subject to wear.
Maintenance intervals
Maintenance must be performed annually in order to ensure the safety and func-
tioning ability of the system.
If you have not concluded a maintenance contract, please notify the Siemens Up-
time Service on time.
CAREWATCH
Display data
At the start of fluoroscopy/acquisition the following system parameters are dis-
played in the lower right area of the live monitor:
1st line
Display of the prefilter in fluoroscopy
2nd line
Display of the prefilter in radiography
With the area dose product meter (optional)
1
3rd line
Area dose product in cGycm²
4th line
During fluoroscopy
Display of the patient entrance dose
2
in mGy/min
In the radiation pauses
the percentage of the patient entrance dose reached related to a config-
urable limit
1
of 0.5 to 5 Gy or
the accumulated patient entrance dose in mGy is displayed.
1
can be configured by the SIEMENS Uptime Service
2
standardized to 30 cm above the tabletop
The measuring device must be calibrated at regular intervals. This is done within
the scope of a maintenance contract. If no maintenance contract have been con-
cluded, the measuring device can be calibrated by the Siemens Uptime Service
of the manufacturer.