System Operation
On-Off/Emergency Stop
AXIOM Iconos R200 AXD3-340.620.13.01.02 7 / 32
In the case of a power failure
The imaging system remains switched on. The generator and table are switched
off. The green lamp on the control unit of the UPS goes out.
The red lamp on the control unit lights up after approx. 4 s.
Now switch the system on within the next 6 s.
If the system is not switched back on within this time, then the imaging system
is shut down in a controlled manner and subsequently restarted.
After restoration of the power supply
The green and red lamps on the control unit of the UPS light up.
Switch the system off.
The imaging system remains switched on.
The generator and table are switched off.
Press the red button on the control unit (Press To toggle Power Supply).
The red lamp on the control unit goes out.
Switch the system on within the next 6 s.
If the system is not switched back on within this time, then the imaging system
is shut down in a controlled manner and subsequently restarted.
Red Emergency STOP button
STOP button on the unit STOP button on the system remote control console
If there is an emergency situation that is potentially hazardous to the patient, the
operator or the system,
Press one of the red stop buttons on your system.
All system movements are stopped;
The generator and the imaging system still remain switched on and ready;
Radiation is interrupted;
fluoroscopy can be switched on again consciously after letting the foot-
switch go and actuating it again;
Injection with the synchronized injector is interrupted.
STOP
AXIOM
STOP
Start
DR
1 2 3
1 2 3
R
R
1,50
1,15
STOP
DR DR DR
STOP