System Operation
Functional and Safety Check
10 / 32 AXD3-340.620.13.01.02 Operator Manual
Run a functional test of the STOP buttons by
performing an arbitrary system movement and
pressing the STOP button during the movement.
The movement concerned must stop immediately.
Then unlock the button (turn it clockwise).
Run a functional test of the proximity switches (collision protection) by
performing an arbitrary system movement and
pressing a proximity switch during the movement.
The movement concerned must stop immediately.
After releasing the proximity switch, all system movements are possible
again.
During the examination
Check the radiation ON indicator. It should light up only
if one of the fluoroscopy switches has been pressed or
for the duration of an X-ray exposure.
Check the patient positioning devices, e.g. use of the grip protection strip and
handgrip.
Switch system movements on only if,
there is no danger of injury to the patient or third parties,
there are no objects blocking the path of system movements.
Monthly tests
Run a functional test of the automatic dose rate control and automatic format
collimation as described in the “Safety” operating instructions.
Legally required tests
Inform yourself in the “Safety” operating instructions.
Warning
If radiation (fluoroscopy, digital radiography) is released and the radiation ON in-
dicator lights up but no fluoroscopic image appears on the monitor (error in the
television chain)
immediately switch the system off using the emergency shutdown button
and
notify the SIEMENS Uptime Service.