AXIOM Iconos R200 AXD3-340.620.13.01.02 11 / 32
System Operation
System Settings
General notes
Collision protection / Safety areas
All system movements stop when a collision protection device is activated.
The movement is stopped when a collision zone is reached.
3D support If the 3D support (optional) is not in its park position, then the system move-
ments are disabled.
This system movement is enabled by simultaneously actuating the bypass key
and an operating element.
Press and hold this key
Start the system movement after 2s.
Operating elements for system positions
Operating elements with deadman functions
All operating elements for the system settings are designed as deadman func-
tion, i.e. system settings are active only as long as the operating element is actu-
ated.
Operating elements with LEDs
LED lights up: function is selected
LED does not light up: function is deselected
LED flashes: selected function is not yet concluded or message about a particular
status (see relevant chapter for description).
P
Warning
Immediately press the nearest STOP button in the case of
danger to the patient.
All system movements are stopped.
After removing the cause of the danger, unlock the STOP button by turning
it to the right.