AXIOM Iconos R200 AXD3-340.620.13.01.02 11 / 32
System Operation
System Settings
General notes
Collision protection / Safety areas
❏ All system movements stop when a collision protection device is activated.
❏ The movement is stopped when a collision zone is reached.
3D support ❏ If the 3D support (optional) is not in its park position, then the system move-
ments are disabled.
This system movement is enabled by simultaneously actuating the bypass key
and an operating element.
◆ Press and hold this key
◆ Start the system movement after 2s.
Operating elements for system positions
❏ Operating elements with deadman functions
All operating elements for the system settings are designed as deadman func-
tion, i.e. system settings are active only as long as the operating element is actu-
ated.
❏ Operating elements with LEDs
LED lights up: function is selected
LED does not light up: function is deselected
LED flashes: selected function is not yet concluded or message about a particular
status (see relevant chapter for description).
Warning
Immediately press the nearest STOP button in the case of
danger to the patient.
All system movements are stopped.
◆ After removing the cause of the danger, unlock the STOP button by turning
it to the right.