System Operation
System Settings
12 / 32 AXD3-340.620.13.01.02 Operator Manual
Object-dependent system settings
The agreement of the actuation direction of the operating element with the
movement of the X-ray system (stand with tube assembly and primary collimator,
image receptor unit with image intensifier and spotfilm device) depends upon the
selected object setting.
Left LED lights up: ”Object setting oriented to the monitor” is selected
Right LED lights up: ”Object setting oriented to the system” is selected
Only possible from the system remote control console for the following move-
ments:
Moving the X-ray system longitudinally
Moving the tabletop longitudinally/transversely
Start positions
Automatically determined system settings are performed with the Start button
depending upon the selected operating mode (tomography, Periscanning,
Peristepping, scanning technique).
LED flashes: the approved system settings are not yet concluded.
LED does not light up: the approved system settings are concluded.
General notes on error messages
The system is constantly monitored internally during operation.
When a fault is detected, the system is blocked and the fault is displayed as error
in the message line of the integrated generator control console and stored.
Cancelling blocking:
Briefly press the exposure release button on the generator ON/OFF console
up to the 1st pressure point and let go again. Don’t press more!
The fault message is cancelled as a rule
However, if an error message is still displayed, there is an error.
Call the Siemens Uptime Service
Start
Further operation of the system may be only under increased attention of the
user.