System Operation
System Settings
AXIOM Iconos R200 AXD3-340.620.13.01.02 15 / 32
Tabletop
Moving the tabletop longitudinally/transversely
System remote
control console
◆ Actuate the coordinate switch
– key on the left/right: the tabletop moves transversely
– key above/below: the tabletop moves longitudinally
◆ Actuate the joystick
– Deflect to the left/right: the tabletop moves transversely
Tableside control
panel
◆ Actuate the coordinate switch
1
– key on the left/right: the tabletop moves longitudinally
– key above/below: the tabletop moves transversely
Travel range ❏ In order to avoid collisions with the floor, the system will retract the tabletop
automatically as needed if you tilt the table up or down.
The X-ray system follows this movement synchronously if the available travel
range is sufficient.
The object always remains centered in this case.
❏ If you have tilted the table more than -30° down or more than +30° up, the
system continuously calculates the maximum possible travel range of the ta-
bletop, depending upon the room height and the system position. You can
move the tabletop only within this range.
AXIOM
1
Display in horizontal table position