System Operation
System Settings
AXIOM Iconos R200 AXD3-340.620.13.01.02 21 / 32
Compression device (optional)
(1) Compression device with cone
Using the compression device
Insert the cone arm.
Refer to the ”Accessories” chapter.
Set an oblique projection angle of less than 30°.
Press a key.
key above: decompression
key below: compression.
For further information refer to “Patient positioning” im Register System
Overview/Protective Measures chapter.
For information about patient rescue in case of a power failure see
the chapter On-Off/Emergency Stop
You can apply a maximum compression force of 155 N or 80 N
1
, adjustable in
15 or 8
1
steps.
Movements are no longer possible at a compression force of 50 N and higher.
The cone carriage moves automatically into the park position (topmost decom-
pression position)
in tomography
in Peristepping, Periscanning, scanning technique
when setting a swivel angle greater than 30°.
(1)
AXIOM
The motor-driven compression device is subject to the examiner’s special duty
to take due care regarding the applied compression forces, especially in the
case of frail, sick and elderly patients.
1
configuration for Japan only