System Operation
System Settings
22 / 32 AXD3-340.620.13.01.02 Operator Manual
Cone arm in park
position
Under the following conditions no shadowing of the collimated format will result
from the cone arm in the park position:
Max. 40 cm high, on the horizontal unit with orthogonal collimation and 150
cm SID.
Max. 35 cm high, on the vertical unit with orthogonal collimation or oblique
projection in all unit positions from115 cm SID.
Reading the compression level
System remote
control console
Read the indication on the display.
Monitor Read the indication at the bottom right on the monitor.
Tomographic equipment (optional)
(1) Mirror
(2) Scale
Setting the fulcrum height
The average fulcrum height (120 mm) is set automatically when the system is
switched on. Fulcrum heights of 10 mm - approx. 300 mm above the tabletop can
be set for cassette tomography and digital tomography.
Press one of the keys:
Adjustment upwards;
Adjustment downwards.
(1)
(2)