System Operation
System Settings
24 / 32 AXD3-340.620.13.01.02 Operator Manual
Manual tube assembly rotation
During the rotational movement radiation release is not blocked.
Outside the 0° rotation position, the automatic format collimation system
switches over to the “Manual” mode (free radiation field setting).
Tube assembly rotation is possible from +90° to -90° with stop positions every
10° and from -90° to -180° with stop positions every 30°.
Remove the compression cone if present.
Pull the handle (1) to the front and hold it firmly.
Turn the X-ray tube assembly to the required position and hold it firmly at rest.
Press the handle towards the rear and allow it to engage.
Check for lock-in by turning the tube assembly slightly to the left and right.
Moving the grid into / out of the beam path
Moving the grid into the beam path
Press this key.
LED lights up: the grid is in the beam path.
Moving the grid out of the beam path
Press this key again.
LED dark: the grid is moved out of the beam path.