System Operation
Setting the Image Geometry
AXIOM Iconos R200 AXD3-340.620.13.01.02 29 / 32
By pressing this key you move all semitransparent wedge filters out of the
beam path (e.g. for resetting the wedge filters).
LED lights up: no wedge filter in the beam path
Moving the semitransparent wedge filters
The movements refer to the display on the monitor and always behave the same
even with image reversal.
Coordinate switch for
moving / turning left
1
wedge filter
moving / turning double wedge filter
Coordinate switch for
moving / turning right
1
wedge filter
moving / turning double wedge filter
CAREPROFILE, collimation without radiation (optional)
The collimation of the rectangular/iris collimators and of the semitransparent fil-
ters (option) can be displayed on the LIH image without radiation.
The collimator position is displayed by a contour line as soon as the position of
the collimator is changed.
Refer to the FLUOROSPOT Compact operating instructions for further details.
CAREPOSITION (optional)
The CAREPOSITION function enables you to position the patient anew without
further fluoroscopy with the aid of the last fluoroscopic image (LIH).
Refer to the FLUOROSPOT Compact operating instructions for further details.
1
in relation to monitor image edge