Examination
Fluoroscopy
AXIOM Iconos R200 AXD3-340.620.13.01.02 11 / 52
Fluoroscopy time limit
Fluoroscopy is switched off automatically and can no longer be switched on after
a fluoroscopy time of 10 minutes
1
.
To continue working
let the fluoroscopy switch go and press it once again.
Fluoroscopy warning signal
An acoustic signal sounds after a fluoroscopy time of maximum 5 minutes
2
.
The signal disappears when you press the key on the Fluorospot Compact.
The time for the acoustic signal can be reset to zero at any time.
Automatic format collimation in fluoroscopy
Automatic format collimation is also switched on automatically when fluoroscopy
is released.
I.e. the fluoroscopic field is never larger than the selected image intensifier for-
mat.
You can limit the fluoroscopic field to the required size with the corresponding
operating elements within this area.
Fluoroscopy programs
All parameters necessary for fluoroscopy are stored in programs.
The following programs can be stored:
Standard fluoroscopy
Supervision
Pulsed fluoroscopy (Carevision)
Roadmap standard fluoroscopy
Roadmap Supervision
Roadmap pulsed fluoroscopy
Refer to the Fluorospot Compact Operator Manual.
1
Can be changed on request by SIEMENS Uptime Service, under compliance with the national
regulations.
2
Can be configured from 1 to 5 minutes by the SIEMENS Uptime Service