Further display functions
I
I
The current compensated in the zero is display-
ed here in pA.
15 Display compensated background current
(zero calibration)
Identifier: n =
1 = Oven temperature (100 to 200 ˚C)
2 = Physical section temperature
(10 to 60 ˚C)
3 = Pump temperature (110 to 190 ˚C)
4 = Pump temperature comparison
(110 to 190 ˚C)
5 = Sampling temperature
(110 to 190 ˚C)
6 = Sample gas probe temperature
(110 to 190 ˚C)
7 = Flame temperature (230 to 300 ˚C)
8 = Temperature of flowmeter
(10 to 80 ˚C)
9 = Temperature of board (10 to 70 ˚C)
10 = Flow (50 to 2000 ml/min according
to analyzer status)
11 = Pressure of sample gas/combustion
air (mbar abs.)
12 = Hydrogen pressure (∼ 1000 mbar
absolute)
13 = Fan current (50 to 200 mA)
14 = FID voltage (200 V)
15 = +15 V
16 = –15 V
17 = +24 V
Identifier: n =
0 = Measured value
1 = Raw measured value (pA)
2 = Analog current
3 = Current following adjustment of
preamplifier
4 = Current following zero adjustment
Terminating this function via CL key sets the
identifier automatically to 0.
Identifier n not
equal to 0
6
CL
n
1
Identifier
n
61 Display measured variables
6
CL
Display
0
Identifier
* 60 Display internal values
Display
51
A-27