6 Cleaning
Hematek
®
3000 System – Instruction Manual – Version 1.3 76
6
3. Remove the middle cover from the system, see Figure 6-1.
The rear
trough and guide rail are completely exposed and easily accessible.
Clean drain troughs and rear
guide rail
4. Wet a cotton swab thoroughly with methanol.
1 Lid
2 Middle cover
Figure 6-1 Removing the middle cover
2
1
NOTICE
Damage to or incorrect positioning of the sensing switch fingers can
result from not following recommended cleaning practices
System can fail and staining of the smeared glass slides can be of
inadequate quality.
Always wipe the platen from right to left.