Hematek
®
3000 System – Instruction Manual – Version 1.3 57
8. Dispose of the disposable, absorbent cloths.
Remove pump tubing 9. Left of the pump assemblies, turn the pump tubing release knob to the
unlock position, see Chapter 4.1.1.
10. Disconnect each tube from its cannula.
11. On the rear of the middle panel, disconnect the tube from the nipple. An
example pump assembly is displayed in Figure 7-1.
12. Push the pump arm handle away from the pump assembly as far as
possible.
13. Hold the pump arm handle in this position and pull the plastic cuff from
the pump arm until the tube is completely removed.
14. Execute step 11 to step 13 for each tube.
15. If you want to prepare the system for non-operation, close the lid,
otherwise skip this step.
1 Tube
2 Nipple
3 Middle panel
4 Pump assembly
5 Pump arm
6 Pump arm handle
7 Plastic cuff
Figure 7-1 Removing the pump tubing, rear view