7 Performing maintenance
Hematek
®
3000 System – Instruction Manual – Version 1.3 77
7
6. Make sure that you push each tube until it is flush against the guard of
the cannula.
Thread tubes through pump arms 7. Left of the pump assemblies, turn the pump tubing release knob to the
unlock position, see Chapter 4.1.1.
8. Extend each new tube to its respective pump assembly (in numerical
order from right to left).
9. Push the pump arm handle away from the pump assembly as far as
possible, see Figure 7-3.
10. Thread the free end of the tube into the hole in the pump arm.
11. Push the tube through the hole in the pump arm until the plastic cuff is
flush against the pump arm.
12. Release the pump arm handle.
Connect tubes to nipples 13. On the rear of the middle panel, connect the tube to its proper nipple.
1 Nipple
2 Middle panel
3 Pump assembly
4 Pump arm
5 Pump arm handle
6 Hole in the pump arm
Figure 7-3 Installing the pump tubing, rear view