Hematek
®
3000 System – Instruction Manual – Version 1.3107
7 Performing maintenance
7
1. Raise the lid.
Empty underplaten tubing 2. Remove the three cannulas from the Hematek Stain Pak.
3. Place the tips of the cannulas on a disposable, absorbent cloth to soak
up fluids.
4. Left of the pump assemblies, press the priming button, see
Chapter 4.1.1.
The green light in the priming button goes on and the
pumps start rotating.
5. Wait until the pump tubes are empty and no more fluid accumulates on
the platen.
6. Press the priming button again.
The green light in the priming button
goes off and the pumps stop rotating.
7. Dispose of the disposable, absorbent cloth.
Remove middle cover 8. On top of the system, at the right side, press the power switch.
The
green power light in the power switch goes off, and the fan and the
conveyor spirals stop.
9. Remove the middle cover from the system, see Figure 7-4.
1 Lid
2 Middle cover
Figure 7-4 Removing the middle cover
2
1