Operating
7.12 Open Channel Monitoring (OCM)
HydroRanger 200 HMI
Operating Instructions, 06/2018, A5E36281317-AC
143
1)
Zero head (2.13.4.5.) (Page 244)
Empty (2.2.4.) (Page 170)
D = pipe or sewer diameter
D/2, point of measurement
2)
Primary Measuring Device (2.13.2.)
Palmer-Bowlus
PMD Dimensions (2.13.5.) (Page 246)
OCM Dimension 1 (2.13.5.1.)
Flume width (D)
Maximum Head (2.13.4.2.) (Page 243)
Maximum Flow (2.13.4.3.) (Page 243)
Zero Head (2.13.4.5.) (Page 244)
Flow Time Units (2.13.4.4.) (Page 244)
● Sized by pipe diameter D
● Flume relief is trapezoidal
● Designed to install directly into pipelines and manholes
● Head is referenced to bottom of the throat, not bottom of the pipe
● For rated flows under free flow conditions, the head is measured at a distance of D/2
upstream from the beginning of the converging section.