Parameter reference
8.6 Setup (2.)
HydroRanger 200 HMI
166 Operating Instructions, 06/2018, A5E36281317-AC
8.6.1.9
Short shot duration (2.1.9.)
Adjusts the duration of the short shot transmit pulse, in milliseconds.
Index
Single-point model
Dual-point model
Global
Transducer
Values
Range: 0.000 to 5.000
Altered by
Transducer (2.1.5.) (Page 164)
Related
Number of long shots (2.1.12.) (Page 167)
Long shot duration (2.1.10.) (Page 166)
Long shot frequency (2.1.8.) (Page 165)
8.6.1.10
Long shot duration (2.1.10.)
Adjusts the duration of the long shot transmit pulse, in milliseconds .
Index
Single-point model
Dual-point model
Global
Transducer
Values
Range: 0.000 to 5.000
Altered by
Transducer (2.1.5.) (Page 164)
Related
Number of short shots (2.1.11.) (Page 166)
Short shot frequency (2.1.7.) (Page 165)
Short shot duration (2.1.9.) (Page 166)
Short shot bias (2.11.2.10.) (Page 225)
Short shot floor (2.11.2.11.) (Page 226)
Short shot range (2.11.2.12.) (Page 226)
8.6.1.11
Number of short shots (2.1.11.)
Sets the number of short shots to be fired (and results averaged) per transmit pulse.
Index
Single-point model
Dual-point model
Global
Transducer
Values
Range: 0 to 100
Preset: 1
Related
Number of long shots (2.1.12.) (Page 167)
Short shot frequency (2.1.7.) (Page 165)
Long shot duration (2.1.10.) (Page 166)
Short shot bias (2.11.2.10.) (Page 225)
Short shot floor (2.11.2.11.) (Page 226)
Short shot range (2.11.2.12.) (Page 226)