Commissioning
6.5 Quick Start Wizards
HydroRanger 200 HMI
70 Operating Instructions, 06/2018, A5E36281317-AC
Response rate
Sets the reaction speed of the device to measurement changes in the target range.
Note
Any changes made to Fill rate/minute (2.3.2.) (Page 173), Empty rate/minute (2.3.3.)
(Page 173), or Response rate (2.3.4.) (Page 174) parameters following the completion of
the wizard will supersede the Response Rate setting.
Response Rate always displays in meters per minute (m/min).
Index
Transducer
Options
Slow (0.1 m/min)
*Medium (1.0 m/min)
Fast (10.0 m/min)
Factory setting
Medium (1.0 m/min)
Use a setting just faster than the maximum filling or emptying rate (whichever is greater).
Slower settings provide higher accuracy, faster settings allow for more rapid level
fluctuations.
Dimension A
The height of the vessel bottom when the bottom is conical, pyramidal, parabolic, spherical,
or flat-sloped.
Dimension A as used in Vessel shape (2.7.2.) (Page 186).
Index
Level
Values
Range: 0.0 ... 99.00 m or equivalent, depending on Units (2.1.1.) (Page 162).
Preset: 0.000
Enter ONE of the following:
Height of the tank bottom, if Vessel Shape (2.7.2.) (Page 186) = Conical Bottom,
Parabolic Bottom, Half Sphere Bottom, or Flat Sloped Bottom
Length of one end section of the tank, if Vessel Shape (2.7.2.) (Page 186) = Parabolic
Ends, in Units (2.1.1.) (Page 162)