Quality measurement
The spine coil is positioned on the patient table.
1 Position phantom holder 26 on the table.
2 Place the 5300-ml plastic bottle inside the phantom holder 27.
3 Position the Contour 24 coil between the phantom holder 26 and
the plastic bottle.
4 Secure the coil on the bottle using the straps.
5 Position the coil with the plastic bottle on the phantom holder 26
so that the coil cable points toward the magnet.
5
Quality measurement 5
MAGNETOM Family |
syngo
MR XA30
Print No. MR-06032G.640.01.02.02
19