Siemens AG Healthcare Sector 11/12 Technical data
28
Power consumption
General classifications
System off
1
1. All data incl. cold head compressor, without cooling
5.0 kW
Stand-by
1
5.7 kW
Ready for measurement
1
13.9 kW
Typical measurement
1
20.7 kW
Protection class I
Components used
• Patient table
• Local coils
• Body coil
• ECG/Pulse module
Type B applied part
Type B or BF applied part
Type B applied part
Type BF applied part
IP protection class according to
IEC 60529
IP X0
Explosion protection The MR system is not intended for opera-
tion in areas prone to explosion (e.g.,
highly flammable mixtures of anaesthesia
gases with air or oxygen or nitrous oxide)
Operating mode Continuous operation with short-term load
Disinfection receptors/system compo-
nents
Disinfectants without alcohol, ether
Degree of safety in the presence of a
flammable anesthetics mixture with air
or with oxygen or with nitrous oxide
No AP or APG category equipment
Mains operated equipment with addi-
tional power sources
None