Operator Responsibilities for Maintenance 13
© Siemens, 2009 M7-000.664.01.04.02 Page 13 of 14 MAGNETOM
03.12 CS SD MR
Tim dockable table Preventive measure to
avoid wear and tear
Check the docking mechanism
Check the hydraulic system
Check/refill hydraulic oil
Check/clean connectors of the docking
station
Lubricate RF connectors
Annually
RF connectors Preventive measure to
avoid wear and tear
Check/clean connectors
Lubricate RF connectors
Annually
Comfort kit (option) Preventive measure to
avoid overheating
Replace air filter Every
2years
Software Preventive measure
against data loss in the
case of hard disk failure
Save dynamic data
Clean up directories
Every
3months
1. According to manufacturer recommendations
2. According to manufacturer recommendations
3. Zero helium boil off rate for typical clinical use, depending on sequence and operating time with running helium compressor.
The system needs to be serviced at regular intervals. Undisturbed magnet cooling for 24 hours and 7 days a week.
4. If the system is not connected to Siemens Remote Service (SRS) the replacement has to be performed according to manu-
facturer recommendations. The replacement is performed “on demand” when the system is connected to Siemens Remote
Service (SRS).
Object or function Reason What is checked: Interval