Safety instructions
A.1-12
0.0
MAGNETOM Family - Safety Information
0.0
The operator should be particularly aware of:
A.1
Adherence to the positioning information for patients (to
avoid current loops and burns)
Careful input of the patient weight/position and orientation
Possible peripheral nerve stimulation, as effect of the First
level controlled operating mode on patients and MR work-
ers
A.1
A.1
WARNING
A.1
MR workers and all personnel who have access to the MR
system are not sufficiently informed!
Personal injury, property damage
Ensure that all personnel (incl. cleaning crews, rescue
personnel, etc.) are regularly informed about the
potential risks inherent in MR systems as well as the rel-
evant safety information (for example, regarding mag-
netic forces).
The exclusion zone and corresponding safety measures
must be observed even when the system is switched
off.