Safety instructions
A.1-20
0.0
MAGNETOM Family - Safety Information
0.0
Maintenance/repair
A.1
Responsibility
A.1
As a supplier, Siemens will not be held responsible for the safety,
reliability, and performance of the system in the following
cases:
A.1
■ Installations, additions, adjustments, modifications, and
repairs to the MR system, or changes to the software that
are not performed by Siemens Service.
■ Assemblies are not replaced with original spare parts.
■ The electrical wiring in the room does not meet the require-
ments of VDE regulation 0100-710 or applicable national
laws.
Siemens is not responsible for potential damage in the event
non-authorized personnel refilling the magnet with helium.
A.1
Assembling and modifications during the actual service life
require evaluation to the requirements of IEC 60601-1.
A.1
A.1
A.1
WARNING
A.1
Unauthorized work on the magnet!
Personal injury, property damage
◆ Only authorized personnel (Siemens Magnet Technol-
ogy or Siemens) may perform work on the magnet.
◆ Do not open or remove safety valves and burst disks of
the helium container.
◆ Do not change the standard configuration.