25
Siemens AG Healthcare Sector 11/12 Technical data
Ambient conditions
Control room
Temperature 15–30 °C (59–86 °F)
Relative air humidity 40 to 60%
Absolute air humidity <11.0 g/kg
Electronics room
Temperature 15–30 °C (59–86 °F)
Relative air humidity 40 to 80%
Absolute air humidity <11.0 g/kg
Examination room
Temperature 18–22 °C (64–72 °F)
Relative air humidity 40 to 60%
Absolute air humidity <11.0 g/kg