EasyManua.ls Logo

Siemens MAGNETOM Verio - 5.4 Patient Preparation

Siemens MAGNETOM Verio
140 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Preparing the patient
CAUTION
Heat development during the examination!
Patient burns
Instruct patients to press the squeeze bulb in case of strong
heat sensations.
WARNING
Use of unapproved fMRI stimulation devices for the given
magnetic field strength!
Injury to patient and operating personnel
Ensure that the stimulation devices are approved for the field
strength of your MR systems. For example, devices approved
for low and medium field systems (0.2 – 1.5T) must not be
used on a 3T system.
Informing the patient
1 Please note the safety instructions.
2 Inform the patient about the possible effects of MR examinations
and the risks associated with the magnetic field.
3 Show the patient how to activate the patient alert by pressing the
squeeze bulb.
4 Ensure that the patient holds the squeeze bulb in his/her hand
during the measurement.
Physiological effects
Due to the presence of alternating electromagnetic fields, patients
may experience various physiological effects during MR
measurements:
5.4
5.4.1
5.5
MR system operation 5
MAGNETOM Avanto, MAGNETOM Verio | syngo MR E11
113
Print No. MR-01501G.621.02.02.02

Table of Contents

Related product manuals