Cleaning system components
WARNING
Flammable cleaning or disinfection agents may cause fire or
explosion!
Injury to patient
Disinfect the system components with commercial
disinfecting materials. However, do not use flammable
solvents such as undiluted alcohol, benzine, or acetone.
Follow the manufacturer’s instructions.
The following table lists classes of active agents that have been tested
and are approved. The detergents listed under “not approved” must
not be used.
Approved
Not approved
Aldehydes
Quaternary compounds
Guanidine derivatives
Peroxide compounds
Pyridine derivatives
Alcohol solution (for example, 70%
isopropyl alcohol and 30% water)
Chloro derivatives
Commercially available cleaning
agents, detergent substances
Alkylamine
Organic acids
Benzine undiluted
Acetone
Phenols
Clean the system and its components with a damp cloth, if not
specified otherwise in the individual subchapters.