The following table lists classes of active agents that have been tested
and are approved. The detergents listed under “not approved” must
not be used.
Approved Not approved
Aldehydes
Quaternary compounds
Guanidine derivatives
Peroxide compounds
Pyridine derivatives
Chloro derivatives
Commercially available cleaning
agents, detergent substances
Alkylamine
Organic acids
Benzine undiluted
Acetone
Alcohol solution (for
example, 70% iso-
propyl alcohol and
30% water)
Phenols
1
Carefully wipe the plugs and connectors with the cloth. Do not
touch the contacts.
2 Wash the straps of the patient table at a temperature of 60°C.
To clean the control panels, do not use cleaners or disinfectants
containing alcohol or petroleum ether (“benzine”).
To clean the squeeze bulb, headphones, and their tubes, do not
use cleaners or disinfectants containing alcohol, petroleum ether
(“benzine”), or solutions containing quaternary compounds.
The LCD monitor of the syngo Acquisition Workplace and the video
display are cleaned in the same way.
1 Clean the LCD monitor/video display at least every two months.
2 Prior to cleaning, switch off the LCD monitor/video display and
disconnect the main power plug. But if the monitor/display is under
hot running conditions, please wait till the monitor/display is
chilled (this can take up to one hour).
3 Clean the monitor/video display using a microfiber cloth.
Cleaning the patient table, the
straps, the plugs and
connectors
Cleaning the control units
Cleaning the squeeze bulb,
headphones, and their tubes
Cleaning the LCD monitor/
video display
Maintenance 6
MAGNETOM Avanto, MAGNETOM Verio | syngo MR E11
127